Back
VDA 19 / ISO 16232

Technical Cleanliness to ISO 16232 - how to verify?

Technical cleanliness defines how much residual contamination is acceptable on a component. This guide explains residual dirt analysis to VDA 19 and ISO 16232 - from extraction through gravimetry and particle counting to setting meaningful limit values.

5 minStand: 2026-07Geprüft: Technical editors
View cleaning equipment
50-1000 µm
typical size classes
mg/1000 cm²
gravimetric value
ISO 16232
international standard
3-step
extraction to decay curve
Inhalt
  1. Basics and standards
  2. Residual dirt analysis
  3. Evaluating results
  4. Setting limit values
  5. Frequently asked questions

What does technical cleanliness to VDA 19 mean?

Technical cleanliness assesses particulate contamination that adheres to functionally relevant components and can cause damage in operation. Testing follows VDA 19.1 (inspection) and the technically identical international standard ISO 16232, complemented by VDA 19.2 for cleanliness in the assembly environment.

Hard particles such as metal chips, burrs or mineral residues are especially critical. In hydraulics, injection technology or electronics, particles as small as 50 µm can block valves, damage seals or cause short circuits.

VDA 19 does not fix limit values; it describes the test method. The specific cleanliness requirements are agreed between manufacturer and supplier on a per-component basis.
  • VDA 19.1 / ISO 16232: residual dirt analysis on the component.
  • VDA 19.2: cleanliness of the manufacturing and assembly environment.
  • Goal: reproducible, comparable results across the supply chain.

How does residual dirt analysis work?

The analysis splits into three steps: extraction, filtration and evaluation. First the particles are released from the component with a test fluid, then captured on an analysis filter, and finally weighed or measured under a microscope.

Extraction efficiency must be validated with a decay curve: extraction is repeated until one stage yields less than 10 percent of the dirt collected before. Only then is the component considered largely free of residual particles.

A blank value (test without a component) is mandatory. It proves that environment, fluid and equipment do not introduce more particles than the component itself releases.

Gravimetry or particle counting - what to measure?

The two evaluations complement each other. Gravimetry gives the total dirt mass in milligrams per component or normalised per 1000 cm² of surface. Particle counting under a light microscope records number and size and assigns them to size classes.

  • Gravimetry: simple figure for total load, ideal for series monitoring.
  • Optical particle counting: size distribution by size classes B to N (5 µm to over 1000 µm).
  • Metallographic analysis: separates shiny metallic from non-metallic particles.
  • Largest single particle: often decisive for functional damage.
Cleanliness is stated as a CCC code (Component Cleanliness Code), for example as particle count per size class. Requirement and result can then be compared directly.
Cleaning equipment

Matching cleaning and inspection gear for the series.

Read the guide

How do you set sensible limit values?

Limit values derive from the component function, not from the standard. What matters is the smallest safety- or function-critical clearance - for example the guide clearance of a valve or the electrode spacing on a circuit board.

A four-step approach works well in practice: determine the critical size, measure a reference series, capture the statistical spread, and set the limit with a safety margin above the real process level. This keeps the value achievable yet protective.

Always define the test procedure too: extraction method, fluid, filter pore size and test area. A limit value without its method is not comparable.

Frequently asked questions

How do VDA 19 and ISO 16232 differ?

Both standards describe the same residual dirt analysis in content. ISO 16232 is the international version, VDA 19.1 the German edition; VDA 19.2 adds cleanliness of the manufacturing environment.

Which extraction methods are permitted?

Spraying, rinsing, ultrasound and agitation are common. The method must suit the component and be validated with a decay curve so that particle release is reproducible.

Does the standard set fixed limit values?

No. VDA 19 and ISO 16232 only govern the test method. The specific cleanliness limits are agreed between customer and supplier per component based on its function.

Why is the blank value so important?

The blank value shows how many particles come from environment, test fluid and equipment. Only when it is clearly below the component value is the measurement reliable.

Verify technical cleanliness reliably?

We supply cleaning and analysis equipment for residual dirt testing to VDA 19 and ISO 16232 - from extraction to the analysis filter.

Standard-compliant

Equipment matched to VDA 19 and ISO 16232.

Measurably clean

Gravimetry and particle counting documentable.

Validatable

Decay curve and blank value reliably implemented.

Expert advice

Specialists support limit values and method.

More guides